Categories Environmental Test Chamber
Brand Name: Sinuo
Model Number: SN8816
Certification: Calibration Certificate (Cost Additional)
Place of Origin: China
MOQ: 1
Payment Terms: T/T
Delivery Time: 60 days
Supply Ability: 10 sets/month
Packaging Details: Plywood case
Price: Customized
Applied standard: IEC60068-2-5, MIL-STD810D, etc.
Light source: Metal halide lamp
Lamp power: 4KW
Lamp lifetime: About 500 hours
Total irradiance: 1120 W/m2±10% W/m2
Filter type: Sunlight filter
Chamber Temperature range: RT+10~70℃
Water Spray Cycle: 18min/102min
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

Solar Radiation Metal Halide Xenon Aging Test Chamber UV IEC 60068-2-5


IEC 60068-2-5 Solar Radiation Metal Halide Lamp Aging Test Chamber


Complied Standards & Clauses


The Solar Radiation Test Chamber is designed and manufactured according to the standard requirement of IEC60068-2-5:2010, MIL-STD-810, DIN 75220 and etc.


Equipment Overview


The test chamber is mainly used to test the aging of automobiles, coatings, rubber, plastics, pigments, adhesives chemicals, fabrics, aerospace, naval vessels, electronics industry, packaging industry, and etc. by exposing the material sample to the light and heat radiation of the metal halide lamp.


It stimulates the full sunlight spectrum of halide lamp to reproduce the destructive light waves under different environment, which focuses on accelerating the photochemical effect of solar radiation. It can be used to detect the impact of long-term solar radiation on equipment, which can provide corresponding environmental simulations and accelerated tests for scientific research, product development and quality control.


This equipment can be used to select new materials, change existing materials, or evaluate changes in durability after changes in material composition. It can well simulate changes in materials exposed to sunlight under different environmental conditions.


Technical Parameters


Internal chamber size

1000×1000×1000mm(W*D*H)

Or can be customized

External chamber sizeApprox. 1620×1600×2250mm(W*D*H) subject to actual size
Spectral energy distributionSpectral range

Ultraviolet

rays A

Ultraviolet

rays B

Visible lightInfraredTotal radiation
Wavelength range300-320nm320-400nm400-800nm800-2450nm300-2450nm
Radioactivity4.06W/ m270.5W/ m2604.2 W/ m2186W/ m21090W/ m2
Percentage0.4%604%55.4%3708%100%
Number of tubesMetal halide lamp: 1PC; Power: 4KW
Total irradiance280~3000nm; 1120 W/m2±10% W/m2
Lamp lifetimeAbout 500 hours
Filter typeSunlight filter
Internal chamber temperature rangeRT+10~70℃
Temperature heating/cooling rateHeating or cooling at a rate of 1℃/min
Black standard temperature range40℃~100℃, adjustable(light cycle);25~85℃(dark cycle)
ThermometerBlack standard thermometer BST
Temperature accuracy±0.1℃
Temperature fluctuation±0.5℃
Temperature variation±3℃
Humidity range30%~98%RH (in light off); 20%~60%RH(in light on)
Humidity accuracy±1%RH
Humidity fluctuation±3%RH
Humidity variation≤±5%RH
Photoperiod time1S~999H59M59S, continuous photoperiod time can be adjustable
Rainfall time1S~9999min, continuous rainfall time can be adjustable
Rainfall cycle1~240min interval (off) rainfall cycle time can be adjustable
Water spray cycle18min/102min or 12min/48min (water spray time/non-water spray time)
Spray pressure0.12~0.15Mpa
Nozzle apertureФ0.8mm
Door sealantSilicone rubber tight sealing device
Sample holder materialSUS304 stainless steel
Internal chamber materialSUS304 stainless steel
External chamber materialCold rolled plate with stoving vanish
Protection SystemOver-temperature protection, leakage, short-circuit, phase-loss protection, buzzer alert when the test time is up
Environmental conditionsTemperature 5℃~28℃; relative temperature ≤85%RH; atmospheric pressure 86kPa~106kPa
Power supplyAC380V±10% (three-phase five-wire, make a 63A air switch separately, the air switch must have wiring terminals, and the power cord must be more than 6 square meters; 50±10%; 16KW).